Home   /   Методики   /   External, reflex techniques

External, reflex techniques

  • on acupuncture points (AP), laser acupuncture
  • on Zakharyin – Ged zones (dermatomes)
  • on paravertebral zones

 

THE REQUIRED EQUIPMENT:

  1. LASMIK laser therapy device
  2. Matrix laser head of pulsed illumination ML-904-80
  3. Laser head of pulsed illumination LO-904-20
  4. Laser head of continuous illumination KLO-635-15
  5. Mirror nozzle ZN-35
  6. Magnetic nozzle ZM-50
  7. Acupuncture nozzle A-3
  8. Transparent nozzle for PMN matrix heads

Table 3

Laser Acupuncture Parameters

Parameter Value Note
Wavelength of laser light, nm (spectrum) 525 (green) On auricular ТА
635 (red) On corporal  ТА
Laser mode Continuous or modulated
Frequency Hz As prescribed Only for modulated mode
Power 0,5–1 525 nm
2–3 635 nm
Exposure for 1 zone, min 5–10 On auricular TA  (AP)
20–40 On corporal  TA  (AP)
Number of impact zones Up to 15
Localization As prescribed On auricular TA  (AP)
As prescribed On corporal  TA  (AP)
Technique contact Through an acupuncture tip
Number of procedures per course 10–12 daily

Note. * – accupuncture tip output

 

Impact on Zakharyin – Ged zones (dermatomes)

Impact parameters are identical to the contact-mirror technique (Table 1). The optimal use of the matrix pulse IR-emitting heads, for example, ML-904-80 for LASMIK series. Wavelength 904 nm, power 40–60 W, frequency 80–150 Hz, exposure 1.5–2 min per zone, contact. It is also possible to use a laser emitting head with a single pulsed IR laser (LO-904-20 for LASMIK series), but contact-mirror technique is required (with a mirror nozzle ZN-35 or ZN-50). The wavelength is 904 nm, the pulse power is 10–15 W, the frequency is 80 Hz, the exposure time is 1.5–2 min per zone. In one session up to 4-6 zones. Variations in power and frequency are not allowed.

 

Impact on the paravertebral zones

The exposure parameters are identical to the contact-mirror method (Table 1), but only the pulsed IR LILI is used, two laser emitting heads with one laser (LO-904-20 for LASMIK series with a mirror nozzle ZN-35, symmetrical. Wavelength 904 nm, pulse power 10–15 W, frequency 80–150 Hz, stable, paravertebral, on the projection of sympathetic nodes, 1 min exposure to one zone. In some techniques, exposure time is allowed to vary with increasing number of fields, sometimes labile techniques are used.