THE REQUIRED EQUIPMENT:

  1. LASMIK laser therapy device
  2. Laser head of pulsed illumination LO-904-20
  3. Laser head of pulsed illumination LO-904-80
  4. Laser head of continuous illumination KLO-635-15
  5. Mirror nozzle ZN-35
  6. Mirror nozzle ZN-50
  7. Magnetic nozzle ZM-50

 

PARAMETERS OF CONTACT AND MIRROR TECHNIQUE OF LASER THERAPY *

Table 1

Parameter Value Note
Wavelength of laser light, nm (spectrum) 445 (blue), 525 (green), 635 (red), 780, 808, 904 (IR) Single laser head
Laser mode Continuous 445, 525, 635, 780, 808 нм
Pulsed 635 and 904 nm
Light pulse duration, ns 100–150 Foe pulse mode
Power 10–40 mW Continuous mode
5–25 W Pulse mode
Power density (more absorption – less value) 5–40 mW/cm2 Continuous mode
5–15 W/cm2 Pulse mode
Frequency Hz 80–150 For pulse mode
Exposure for 1 zone, min 2 or 5
Number of impact zones 1–4
Localization On the affected area
Technique Contact-mirror * With the use of mirror and magnetic nozzles
Number of procedures per course 5–12 Daily or every other day

Note. * – parameters of the distant technique are identical, but the laser emitting head is located at a distance from the body surface

 

 

Parameters of the contact technique of LASER THERAPY

Table 2

Parameter Value Note
Wavelength of laser light, nm (spectrum) 780, 808, 904 (IR) Single laser head
Laser mode Continuous 780 and 808 nm
Pulse 904 nm
Light pulse duration, ns 100–150 For pulse mode
Power 100–200 mW 780 and 808 nm
80–100 W 904 nm
Power density (more absorption – less value) Maximum
Frequency Hz 3000–10 000 For pulse mode
Exposure for 1 zone, min 5 In a number of techniques allowed up to 30 minutes
Number of impact zones 1–4 Most often symmetrical
Localization On affected area
Technique Contact Directly by laser diode
Number of procedures per course 15–20 Daily. The course is repeated after 1 month.