External, reflex techniques

  • on acupuncture points (AP), laser acupuncture
  • on Zakharyin – Ged zones (dermatomes)
  • on paravertebral zones

 

THE REQUIRED EQUIPMENT:

  1. LASMIK laser therapy device
  2. Matrix laser head of pulsed illumination ML-904-80
  3. Laser head of pulsed illumination LO-904-20
  4. Laser head of continuous illumination KLO-635-15
  5. Mirror nozzle ZN-35
  6. Magnetic nozzle ZM-50
  7. Acupuncture nozzle A-3
  8. Transparent nozzle for PMN matrix heads

Table 3

Laser Acupuncture Parameters

Parameter Value Note
Wavelength of laser light, nm (spectrum) 525 (green) On auricular ТА
635 (red) On corporal  ТА
Laser mode Continuous or modulated –
Frequency Hz As prescribed Only for modulated mode
Power 0,5–1 525 nm
2–3 635 nm
Exposure for 1 zone, min 5–10 On auricular TA  (AP)
20–40 On corporal  TA  (AP)
Number of impact zones Up to 15 –
Localization As prescribed On auricular TA  (AP)
As prescribed On corporal  TA  (AP)
Technique contact Through an acupuncture tip
Number of procedures per course 10–12 daily

Note. * – accupuncture tip output

 

Impact on Zakharyin – Ged zones (dermatomes)

Impact parameters are identical to the contact-mirror technique (Table 1). The optimal use of the matrix pulse IR-emitting heads, for example, ML-904-80 for LASMIK series. Wavelength 904 nm, power 40–60 W, frequency 80–150 Hz, exposure 1.5–2 min per zone, contact. It is also possible to use a laser emitting head with a single pulsed IR laser (LO-904-20 for LASMIK series), but contact-mirror technique is required (with a mirror nozzle ZN-35 or ZN-50). The wavelength is 904 nm, the pulse power is 10–15 W, the frequency is 80 Hz, the exposure time is 1.5–2 min per zone. In one session up to 4-6 zones. Variations in power and frequency are not allowed.

 

Impact on the paravertebral zones

The exposure parameters are identical to the contact-mirror method (Table 1), but only the pulsed IR LILI is used, two laser emitting heads with one laser (LO-904-20 for LASMIK series with a mirror nozzle ZN-35, symmetrical. Wavelength 904 nm, pulse power 10–15 W, frequency 80–150 Hz, stable, paravertebral, on the projection of sympathetic nodes, 1 min exposure to one zone. In some techniques, exposure time is allowed to vary with increasing number of fields, sometimes labile techniques are used.